Not to scale. Lens shown enlarged for detail.
Acriva UD and Acriva BB are indicated for the primary implantation of a posterior chamber monofocal intraocular lens in adult patients undergoing cataract surgery or clear lens extraction.
Adult patients undergoing cataract surgery or clear lens extraction, at one chosen focal distance.
Best corrected vision at the chosen distance. Spectacles should be expected for the remainder of the range, and the target refraction agreed with the patient before surgery.
Significant corneal pathology, uncontrolled glaucoma, or inadequate capsular support.
Clinical suitability is judged case by case by the prescribing physician.
All three models share a 6.00 mm aspheric optic in hydrophilic acrylic with a hydrophobic surface. Acriva UD is available as UD 613 (Modified C). Acriva BB is available as BB UD 613 (Modified C) and BB UDM 611 (plate).
All powers in 0.50 D increments. Supplied non-preloaded.
Acriva UD and Acriva BB are supplied non-preloaded and loaded at the time of surgery.
Fill the cartridge with Protectalon OVD and seat the lens in the haptic grooves.
AcriJET FLY 2.2 delivers Acriva UD 613, Acriva BB UD 613 and Acriva BB UDM 611 up to 32.00 D.
Optical constants are starting values. Surgeons should optimise them with their own outcomes.
For healthcare professionals only. Read the instructions for use before use. Product availability varies by country.
Acriva UD has a UV filter. Acriva BB adds a natural chromophore at 0.02% concentration that also filters violet and blue light.
613 models have a Modified C haptic and a 13.00 mm overall diameter. 611 models have a plate haptic and an 11.00 mm overall diameter.
No. All three models are non-preloaded IOLs. They are loaded into an injector at the time of surgery.
The nominal acoustic A constant is 118.0. Optical constants are SRK/T 118.40, SRK-II 118.40, Haigis a0 0.91, a1 0.40, a2 0.10, Hoffer Q pACD 5.20, Holladay 1 SF 1.45 and Barrett LF 1.57. Acriva BB UD 613 uses the same values.
The nominal acoustic A constant is 118.0. Optical constants are SRK/T 118.70, SRK-II 118.70, Haigis a0 1.11, a1 0.40, a2 0.10, Hoffer Q pACD 5.37, Holladay 1 SF 1.62 and Barrett LF 1.73.
Plate haptic models (611) cover 0.0 D to +45.00 D. Modified C models (613) cover 0.0 D to +32.00 D as standard, with -20.00 D to -0.50 D and +32.50 D to +45.00 D on special production. All in 0.50 D increments.
The Acriva Trinova Pro C and Acriva BB Toric share the same Acriva UD platform. Select by indication.