Acriva UD packaging beside the intraocular lens

Acriva UD and Acriva BB:
Monofocal IOLs

Monofocal Ultra Definition Non-preloaded CE (MDD)

Not to scale. Lens shown enlarged for detail.

Indication

Primary cataract: single focal distance

Acriva UD and Acriva BB are indicated for the primary implantation of a posterior chamber monofocal intraocular lens in adult patients undergoing cataract surgery or clear lens extraction.

01 Indicated for

Adult patients undergoing cataract surgery or clear lens extraction, at one chosen focal distance.

02 What to expect

Best corrected vision at the chosen distance. Spectacles should be expected for the remainder of the range, and the target refraction agreed with the patient before surgery.

03 Contraindicated in

Significant corneal pathology, uncontrolled glaucoma, or inadequate capsular support.

04 Assessed individually

Clinical suitability is judged case by case by the prescribing physician.

Acriva UD packaging
Acriva UD 613 monofocal IOL with Modified C haptic
Acriva BB UD 613 blue light filter IOL with Modified C haptic
Acriva BB UDM 611 blue light filter IOL with plate haptic
The Platform

One aspheric optic. Two haptic designs.

All three models share a 6.00 mm aspheric optic in hydrophilic acrylic with a hydrophobic surface. Acriva UD is available as UD 613 (Modified C). Acriva BB is available as BB UD 613 (Modified C) and BB UDM 611 (plate).

01
Acriva UD: clearUV filter. Model UD 613.
02
Acriva BB: blue light filterA natural chromophore at 0.02% filters UV, violet and blue light. Models BB UD 613 and BB UDM 611.
03
Two haptics613: Modified C, 13.00 mm, designed for long-term centration. 611: plate, 11.00 mm, suitable for micro-incision surgery.
The Acriva Family
Models

Two filters, two haptics

All powers in 0.50 D increments. Supplied non-preloaded.

Optical Parameters

Technical specifications

Material
Hydrophilic acrylic with hydrophobic surface
Optic Design
Monofocal, aspheric (-0.165 µm), 360° enhanced square edge
Optic Diameter
6.00 mm
Overall Diameter
13.00 mm (613) · 11.00 mm (611)
Refractive Index
1.462 ± 0.002 (wet) · Abbe 58
Power Tolerance
± 0.25 D
A-Constant (613)
Acoustic 118.0 · SRK-II 118.40 · SRK/T 118.40 · Haigis 0.91 / 0.40 / 0.10 · Hoffer Q pACD 5.20 · Holladay 1 SF 1.45 · Barrett LF 1.57
A-Constant (BB UDM 611)
Acoustic 118.0 · SRK-II 118.70 · SRK/T 118.70 · Haigis 1.11 / 0.40 / 0.10 · Hoffer Q pACD 5.37 · Holladay 1 SF 1.62 · Barrett LF 1.73
Delivery System
AcriJET FLY 2.2 (non-preloaded)
Haptic Angulation
0°
Surgical Technique

AcriJET FLY 2.2 delivery

Acriva UD and Acriva BB are supplied non-preloaded and loaded at the time of surgery.

01 Load with Protectalon

Fill the cartridge with Protectalon OVD and seat the lens in the haptic grooves.

02 Inject through 2.2 mm

AcriJET FLY 2.2 delivers Acriva UD 613, Acriva BB UD 613 and Acriva BB UDM 611 up to 32.00 D.

03 Optimise the constants

Optical constants are starting values. Surgeons should optimise them with their own outcomes.

FAQ

Frequently asked questions

For healthcare professionals only. Read the instructions for use before use. Product availability varies by country.

What is the difference between Acriva UD and Acriva BB?

Acriva UD has a UV filter. Acriva BB adds a natural chromophore at 0.02% concentration that also filters violet and blue light.

What is the difference between the 613 and 611 models?

613 models have a Modified C haptic and a 13.00 mm overall diameter. 611 models have a plate haptic and an 11.00 mm overall diameter.

Are Acriva UD and Acriva BB preloaded?

No. All three models are non-preloaded IOLs. They are loaded into an injector at the time of surgery.

What is the A constant of Acriva UD 613?

The nominal acoustic A constant is 118.0. Optical constants are SRK/T 118.40, SRK-II 118.40, Haigis a0 0.91, a1 0.40, a2 0.10, Hoffer Q pACD 5.20, Holladay 1 SF 1.45 and Barrett LF 1.57. Acriva BB UD 613 uses the same values.

What is the A constant of Acriva BB UDM 611?

The nominal acoustic A constant is 118.0. Optical constants are SRK/T 118.70, SRK-II 118.70, Haigis a0 1.11, a1 0.40, a2 0.10, Hoffer Q pACD 5.37, Holladay 1 SF 1.62 and Barrett LF 1.73.

What power range is available?

Plate haptic models (611) cover 0.0 D to +45.00 D. Modified C models (613) cover 0.0 D to +32.00 D as standard, with -20.00 D to -0.50 D and +32.50 D to +45.00 D on special production. All in 0.50 D increments.

Acriva IOL Family

Complementary lenses

The Acriva Trinova Pro C and Acriva BB Toric share the same Acriva UD platform. Select by indication.