Enova Advanced Toric packaging beside the intraocular lens

Enova Advanced Toric:
Continuous range, astigmatism corrected

Not to scale. Lens shown enlarged for detail.

Indication

Cataract, presbyopia and corneal astigmatism

The Enova Advanced Toric is indicated for the primary implantation of a posterior chamber extended depth of focus intraocular lens in adult patients with cataract or clear lens extraction who present with regular corneal astigmatism and seek a continuous range of vision from distance through intermediate.

01 Indicated for

Adult patients with regular corneal astigmatism seeking a continuous range of vision from distance through intermediate.

02 Before surgery

Axis alignment is central to the astigmatic outcome and must be planned pre-operatively. Patients should also be counselled that spectacles may still be required for sustained fine near work.

03 Contraindicated in

Irregular or unstable corneal astigmatism, macular degeneration, significant corneal pathology, uncontrolled glaucoma, or pupil abnormalities.

04 Assessed individually

Clinical suitability is judged case by case by the prescribing physician.

Enova Advanced Toric packaging
Enova Advanced Toric intraocular lens
The Platform

Glistening-free hydrophobic acrylic

Every lens in the Enova family is built on the same proprietary hydrophobic acrylic. Glistening, the fluid microvacuoles that form inside the optic over the years and scatter incoming light, is the failure mode this material was designed to eliminate. In an in vitro study at the Intermountain Ocular Research Center, University of Utah, the material showed no glistening formation.

01
No glistening formationShown in an in vitro study at the Intermountain Ocular Research Center, University of Utah.
02
Preloaded Rotaryjet deliverySingle-use injector through a 2.2mm incision. No cartridge preparation, no lens handling.
03
Toric CalculatorPre-operative cylinder and axis planning, with the surgical plan prepared before the patient reaches theatre.
Full Science & Technology
Optical Parameters

Technical specifications

Material
Single-piece 100% Glistening-Free Hydrophobic Acrylic
Optic Design
Extended depth of focus with toric surface: biconvex, aspheric
Cylinder Range
PAD2–PAD6 (+1.00 to +3.75 D)
Overall Diameter
13.00 mm
Optic Diameter
6.00 mm
A-Constant
118.0 (SRK-T 118.7)
Power Range
+10.00 to +30.00 D (0.50 D increments)
Delivery System
Rotaryjet: preloaded, single-use
Incision Size
2.2–2.4 mm
Haptic Angulation
0° (C-loop)
Regulatory
CE (MDD 93/42/EEC) · MDR 2017/745 transition in progress · ISO 13485:2016 · MDSAP
Surgical Technique

Rotaryjet insertion: Enova Advanced Toric

The Enova Advanced Toric is supplied preloaded in the Rotaryjet single-use injector. The rotary mechanism advances the lens through a 2.2mm incision with a single smooth rotation, no cartridge preparation, no lens handling, consistent delivery every time.

01
Load confirmation

Inspect the preloaded injector. Confirm lens orientation and injector integrity before use.

02
Incision and insertion

Place the injector tip through the 2.2mm incision. Rotate smoothly, continuous motion, constant pressure.

03
Haptic deployment, centration and axis alignment

Confirm haptic deployment in the capsular bag. Centre the optic and rotate to the planned cylinder axis. Standard aspiration to complete.

Rotaryjet Insertion Technique
2:38 · VSY Biotechnology
Enova IOL Family

Complementary lenses

The Enova Maestro, Advanced EDOF, and Maestro Toric share the same material and delivery platform. Select by indication.